Site Information

 Loading... Please wait...
Need Assistance?   866-451-2257
  • My Account

Optris CTlaser

Pricing starting at $850, for formal quote complete application sheet below
NOTICE: To receive a quote with excellent discounted pricing please call our team at (866) 451-2257 or fill out our online Application Sheet here:

Optris Application Sheet

SKU:
OPTCTL-
Weight:
5.00 LBS
Qty:

Optris CTlaser

Precise focussing and noncontact temperature measurement

The infrared thermometer optris CTlaser LT allows for exact temperature measurement on smallest objects from 0,9 mm at a distance of 70 mm. The IR thermometer optris CTlaser LTF is due to its short configuration time of only 9 ms perfectly suited for surveillance of quick working processes.

Both models offer a robust and premium-quality stainless steel measuring head with an innovative double laser sighting for an exact measuring field marking in every distance.

Key Features:

  • Temperature range: -50°C to 975°C (-58 °F to 1787 °F)
  • Spectral range: 8 to 14 µm
  • Response time: 9 ms (CTlaser LTF) 120 ms (CTlaser LT)
FeatureDetails
Environmental rating

IP 65 (NEMA-4)

Optical resolution

75:1 CTlaser
50:1 CTlaser F

Spectral range 8 – 14 μm
Temperature ranges

–50 °C ... 975 °C
(–58 °F ... 1787 °F)

Temperature resolution (display) 0.1 K / 0.5 K with CTlaser F
Output / analog

Channel 1: 0 / 4 – 20 mA, 0 – 5/ 10 V, thermocouple J, K
Channel 2: sensing head temperature
(– 40 °C ... 85 °C [–40 °F ... 185 °F) as
(0 – 5 V or 0 – 10 V), alarm output

Output / digital

USB, RS232, RS485, CAN, Profibus DP, Ethernet
(optional)

System accuracy2)
(at ambient temp. 23 ±5 °C)
(at ambient tem. 73 ±41 °F)

±1 % or ±1 °C2),3) (CTlaser)
±1.5 % or ±1.5 °C2),3) (CTlaser F)
(±1 % or ±1.8 °F2),3) [CTlaser])
(±1.5 % or ±2.7 °F2),3) [CTlaser F])

Repeatability
(at ambient temp.23 ±5 °C)
(23 ±0.5 °F)

±0.5 % or ±0.5 °C1),2) (CTlaser)
±1 % or ± 1 °C1),2) (CTlaser F)
(±0.5 % or ±0.9 °F)1),2) (CTlaser)
(±1 % or ±1.8 °F)1),2) (CTlaser F)

Power supply 8 – 36 V DC

Signal processing (parameter
adjustable via software)

Peak hold, valley hold, average; extended hold
function with threshold and hysteresis

Inputs

Programmable functional inputs for external
emissivity adjustment, ambient temperature
compensation, trigger (reset of hold functions)

Ambient temperature (TAmb)

–20 °C to 85 °C (–4 °F to 185 °F)
(50 °C (122 °F) with laser ON)

Weight

600 g (21.2 oz) (sensing head)
420 g (14.8 oz) (electronics)

Shock / vibration
IEC 68-2-27: 50 G, 11 ms, any axis / IEC 68-2-6: 3 G, 11 – 200 Hz, any axis
Cable length

3 m (standard), 8 m, 15 m
(9.8 ft [standard], 26.2 ft, 49.2 ft)

Power supply 8 – 36 V DC

1) Different spotsizes for CTlaser F (D:S = 50:1)
2) Whichever is greater
3) At object temperatures >0 °C, ε = 1